Abbott Architect i2000sr
Abbott Immunology Analyzers Refurbished

Abbott Architect i2000sr

by Abbott  ·  Category: Immunology Analyzers

The ARCHITECT i 2000SR immunoassay analyzer meets your laboratory’s high standards by delivering STAT results when you need them. The flexible protocols built into the ARCHITECT i 2000SR enhance laboratory workflow and allow you to report results with confidence.

Product details

All units are refurbished, QC-tested and ready for installation. Specifications and feature information as provided by the manufacturer.

The ARCHITECT i2000SR immunoassay analyzer meets your laboratory’s high standards by delivering STAT results when you need them. The flexible protocols built into the ARCHITECT i2000SR enhance laboratory workflow and allow you to report results with confidence. 

The ARCHITECT i2000SR immunoassay analyzer offers a maximum throughput of up to 200 tests per hour. Featuring a load-up capacity of 135 samples with 35 priority and 100 routine areas, the ARCHITECT i2000SR has 25 refrigerated reagent positions.

ARCHITECT i2000SR can be integrated with the ARCHITECT c8000 or ARCHITECT c16000 to consolidate clinical chemistry and immunoassay on one platform. Two i2000SRs can be connected to form a high-speed i4000SR.

DESCRIPTION 
Methods  CHEMIFLEX
Maximum Throughput  Up to 200 tests/hour
Sample Types Serum, Plasma, Whole Blood, Urine
Sample Tubes

Height: 72-102 mm

Diameter: 9.6-16.1 mm

Sample Cup Yes (50 µL dead volume) 
Sample Capacity 135
Sample Barcode Types Code 39, Codabar, Interleaved 2 of 5, Code 128
Sample Result Storage 50,000
Sample Volume 10-150 μL, Average: 57 μL
Automatic Dilution Yes
Sample Probe Carryover ≤0.1 parts per million
Reagent Capacity 25 refrigerated positions
Reagent Type 100% liquid ready-to-use
Reagent Onboard Stability 14-30 days
Calibration Frequency Calibrate with new lot number, if controls are out of range or if specified otherwise within the package insert
Sample, Clot and Bubble Detection  Yes
Reagent Pressure Monitoring  Yes
Sample Interference Measurement No
System Control Center 1 SCC, with color touchscreen monitor, keyboard and mouse 
Onboard Maintenance Records Yes
Online Error Code Help Yes
Host Interface Bidirectional, serial RS-232 interface, host query option available
Remote Diagnostics AbbottLink
Dimensions
(H x W x D)

48” x 61” x 49”

121.9 x 154.9 x 124.5 cm 

Weight

1,081 lbs

490.3 kg

Electrical Requirements AC 180-264V, 47-63 Hz
Water Requirements Purified water to dilute buffer concentrate
Heat Output* 4280 BTU/hr, running mode 
Sample Loading RSH
Sizing this for your lab?

Tell us your workload — we'll confirm availability, configuration, and lead time for the Abbott Architect i2000sr.